Confiabilidad en Instrumentación

 

 

Back to Cameras

C61 & R91

Orbiter

C61 and R91
 Gamma Cameras

PDF Brochure

PPS Presentation

 

 

Refurbished Gamma CAMERAS FOR USING WITH IM512P/E

  • The old computer and console built-in field correctors are replaced by a last generation PC.

  • All field corrections and image processing are digitally done by the software at high speed. The increase in processing speed is equivalent to moving from a PC-XT at 4.77 MHz to a current PC with multi-core processor at 2GHz or more.

  • Only 2 of the 25 original boards are left. This reduces in more than 10 times the risk of failures due to components.

  • Real time digital energy correction. Pulse height analyzer. Triple window and triple isotope. Isotope data base with peak energy and window width preset by the user.

  • Real time digital linearity correction. Better than 0.4 millimeter.

  • Real time digital field uniformity correction. Each isotope-collimator combination may have its own correction map. Digital iris. Uniformity quality control protocol.

  • Static, Whole Body, Dynamic, Planar Gated, SPECT and Gated SPECT acquisition. Digital persistence. ECG acquisition(analog tracing and gating).

  • Simultaneous acquisition and processing.

  • Acquisition matrices: 64x64, 128x128, 256x256, 512x512, 1024x1024, byte or word.

  • Acquisition zoom: 1, 1.25, 1.33, 1.5, 1.75, 2, 2.5, 3, 3.5, 4, 5, 6 and 7. Digital acquisition pan to center any point of de image by clicking on it.

  • Real time center of rotation correction for X and Y axws. COR quality control protocol.

  • Gamma camera calibration and diagnostic protocol.